Ring Pump RP-H series

RP-H Photo
Download
Catalog (PDF)
Outline drawing (PDF)
3D CAD data (STEP)
2D CAD data (DXF)

  • Pump head replacement type capable of discharging up to 10mL/min.
  • Pump head replacement is a one-touch method using strong magnets.
  • The standard model is compatible with three types of motors: brushed DC motors, brushless DC motors, and stepper motors.
  • The pump uses a brushless DC motor and a stepper motor achieved 5,000 hours∗1 of continuous operation by using needle roller bearings.
       ∗1: It is our experimental values using
    water, and do not guarantee the
    pump lifespan.
  • The pump head can be sterilized.
    • Autoclave (25 minutes at 121℃) 10 times
    is possible. However, pumps using
    "Transmaster TM-15" tube can not be
    autoclaved.
    • Gamma ray sterilization (25kGy) is
  • possible.


Standard model (Spec 1)
Type of drive motor
Pump Model No.
Volt-
age
(DC
 V)
Dis-
charge rate
(mL/
  min)
Tube Pump
head
model
No.
   ∗4
Material
ID x OD
(mm)
Brushed DC motor
RP-H02S-10Z-DC24V
24 3 - 10 Silicone 2.0 x 4.0 RK-
H02S
RP-H02F-10Z-DC24V
24 3 - 10 PharMed
BPT
2.0 x 4.0 RK-
H02F
RP-H1.6S-7Z-DC24V
24 2 - 7 Silicone 1.6 x 3.2 RK-
H1.6S
RP-H1.6F-7Z-DC24V
24 2 - 7 PharMed
BPT
1.6 x 3.2 RK-
H1.6F
RP-H1.6R-7Z-DC24V
24 2 - 7 Versilon
F-5500-A
1.6 x 3.2 RK-
H1.6R
RP-H1.6W-7Z-DC24V
24 2 - 7 Trans-
master
TM-15
1.6 x 3.2 RK-
H1.6W
Brushless DC motor∗2
RP-H02S-10Z-DC24VL
24 3 - 10 Silicone 2.0 x 4.0 RK-
H02S
RP-H02F-10Z-DC24VL
24 3 - 10 PharMed
BPT
2.0 x 4.0 RK-
H02F
RP-H1.6S-7Z-DC24VL
24 2 - 7 Silicone 1.6 x 3.2 RK-
H1.6S
RP-H1.6F-7Z-DC24VL
24 2 - 7 PharMed
BPT
1.6 x 3.2 RK-
H1.6F
RP-H1.6R-7Z-DC24VL
24 2 - 7 Versilon
F-5500-A
1.6 x 3.2 RK-
H1.6R
RP-H1.6W-7Z-DC24VL
24 2 - 7 Trans-
master
TM-15
1.6 x 3.2 RK-
H1.6W
Stepper motor∗3
RP-H02S-10Z-DC5VS
5 0.03 - 10 Silicone 2.0 x 4.0 RK-
H02S
RP-H02F-10Z-DC5VS
5 0.03 - 10 PharMed
BPT
2.0 x 4.0 RK-
H02F
RP-H1.6S-7Z-DC5VS
5 0.02 - 7 Silicone 1.6 x 3.2 RK-
H1.6S
RP-H1.6F-7Z-DC5VS
5 0.02 - 7 PharMed
BPT
1.6 x 3.2 RK-
H1.6F
RP-H1.6R-7Z-DC5VS
5 0.02 - 7 Versilon
F-5500-A
1.6 x 3.2 RK-
H1.6R
RP-H1.6W-7Z-DC5VS
5 0.02 - 7 Trans-
master
TM-15
1.6 x 3.2 RK-
H1.6W
∗2. The brushless DC motor have a built-in
driver board
. The pump also comes with a
500mm long wire with a connector for
connecting to the driver board.
∗3. A motor controller is necessary to drive
the stepper motor. (A dedicated motor
controller (RE-C700) sold separately is
also available.)
A 300mm long connection wire comes
out from the stepper motor. The wiring
does not have a connector for connecting
RE-C700.
The RE-C700 can be used with wiring
without a connector, but we can also
support wiring with a connector that
makes it easy to connect to the RE-C700,
so please contact us if you require it.
∗4. Some of Saint-Gobain's tube names are
changed a few years ago.
Versilon F-5500-A

Fluran F-5500-A (Old name)

Models using the DC brush motor
Outline drawing
RP-G3 Outline drawing
Mounting holes
RP-G3 Mounting hole drawing
*Do not install the lead wire
in the downward direction.
Discharge direction
RP-G3 Discharge direction
  • Discharge direction when the positive side of DC
    power is connected to the terminal with the + mark
    and the negative side is connected to the terminal
    on the opposite side.


Models using the DC brushless motor
Outline drawing
RP-G3 Outline drawing
Mounting holes
RP-G3 Mounting hole drawing
*Do not install the lead wire
in the downward direction.
Discharge direction
RP-G3 Discharge direction
  • Discharge direction when
    the rotation direction is
    CW.


Models using the stepper motor
Outline drawing
RP-G3 Outline drawing
Mounting holes
RP-G3 Mounting hole drawing
*Do not install the lead wire
in the downward direction.
Discharge direction
RP-G3 Discharge direction
  • Discharge direction when
    the rotation direction is
    CW.


Driver board for brushless DC motor

The brushless DC motor used in the RP-H series has the built-in driver board.
The driver has the following functions: speed control, rotation direction switching, braking, and rotation signal output of 180 pulses per pump revolution.∗4
The pump is supplied with wiring with connectors for external connections.
The driver input/output details are as follows.

RP-H ブラシレス接続

∗4 & 5 : Since the motor outputs 6 pulses per revolution and the reduction gear ratio
is 1/30, the number of pulses per
revolution of the pump is 6 x 30 = 180.

RP-H ブラシレス接続図